Asked by:Nina8310 2026-07-04 Views 272 4 answers
Back to List ✏️ Ask a Question
💡 How to Answer This Question
  1. 1
    Read the Context

    Read the full description, category and tags to understand the real scenario.

  2. 2
    Give an Actionable Answer

    Give concrete steps, cautions and reasoning — not just a conclusion.

  3. 3
    Cite Sources

    Reference standards, papers or product manuals to make the answer more credible.

  4. 4
    Wait for Acceptance

    You will be notified if accepted. Continue the discussion in comments.

Tags:

All Answers4
T
TonyAssay3117 2026-09-20
The composition of the lysis buffer is critical. Avoid using RIPA containing strong detergents such as SDS, NP-40, or sodium deoxycholate, as these can disrupt antigen-antibody binding, resulting in high background or false positives. Instead, use PBS or a hypotonic buffer containing 0.1% Triton X-100, supplemented with protease inhibitor cocktail and PMSF. Lyse on ice, then centrifuge at 4°C and collect the supernatant. Inhibitors generally do not interfere with ELISA, but note that PMSF has a short half-life and should be added fresh immediately before use.
W
WendyAssay 2026-09-13
High background in brain tissue is mostly due to matrix effects. Do not measure the original sample directly at first; use the kit sample diluent to prepare three dilution levels of 1:2, 1:5, and 1:10 and back-calculate. If the back-calculated concentration increases with the dilution factor, it indicates matrix inhibition or false signal; if the back-calculated concentration decreases, it may be background interference. Hippocampal samples have high protein content, so it is recommended to start testing at at least 1:5. If the sample volume is limited, first take 3 uL for a dilution pilot test; there is no need to use more than ten microliters at once.
T
Tube6470 2026-09-21
For low sample volumes, you can look for high-sensitivity or low-sample-volume versions of kits, which typically allow 1:1 or 1:2 dilution to reduce sample consumption. A dozen or so microliters is enough for one replicate well, but it is recommended to reserve at least enough for upper and lower sample replicates, so prioritize kits with a standard curve sensitivity of 1-2 pg/mL. In terms of brands, you can look into Xinbosheng, Lianke, Thermo high-sensitivity series, etc., but be sure to confirm whether they have been validated for brain tissue homogenate, and do not rely solely on serum data.
董
董阳 2026-09-20
When troubleshooting background, set up a sample blank: add only the sample and diluent to the test wells, without detection antibody/HRP-conjugated antibody. If the color still develops strongly, it indicates endogenous peroxidase or a reaction between sample components and the substrate; switch to a diluent containing HRP inhibitor or purchase a kit specifically designed for tissue homogenates. For plate washing, it is recommended to use a plate washer or manually wash 5 times with 300 µL per well. Regarding lead time, domestic in-stock products are generally 2–5 days, while imported high-sensitivity products may take 1–2 weeks; confirm before placing an order.
Write an AnswerSign in to answer