Welcome to BioReagent Hub Hotline:19920083827
Question Detail

For the pretreatment of animal serum drug residues, chloroform is used for degreasing. Recommendations for corrosion-resistant consumables are requested.

Compliance
Asked by:Elisa2658 2026-03-10 Views 211 4 answers
Back to List ✏️ Ask a Question
💡 How to Answer This Question
  1. 1
    Read the Context

    Read the full description, category and tags to understand the real scenario.

  2. 2
    Give an Actionable Answer

    Give concrete steps, cautions and reasoning — not just a conclusion.

  3. 3
    Cite Sources

    Reference standards, papers or product manuals to make the answer more credible.

  4. 4
    Wait for Acceptance

    You will be notified if accepted. Continue the discussion in comments.

Tags:

All Answers4
E
Elisaer 2026-09-13
耐腐蚀耗材除了看材质,还要看批次稳定性。建议固定一家能提供COC/COA的耗材供应商,到货后抽几支做氯仿浸泡和离心测试,确认盖垫无溶胀、无漏液。枪头如果发现吸完氯仿后内壁发白或移液体积异常,就说明材质不耐受,应停止使用。买规格方面,我们按一个月使用量备货,2ml管和15ml管各备一箱,别一次囤太多以免塑料老化。具体价格按品牌、是否无菌、是否带PTFE垫差异较大,以供应商报价为准。
王
王阳刚 2026-09-21
氯仿对天然橡胶、硅胶垫确实会溶胀,所以离心管盖垫要选PTFE/硅胶复合垫,PTFE这一面接触溶剂,硅胶提供弹性密封。管体首选PP(聚丙烯),不要用PC(聚碳酸酯)或普通PS,氯仿会让它们发雾或裂。枪头也选PP材质,如果有条件用带滤芯的,避免氯仿蒸气反冲腐蚀移液器。我们做药残前处理用2ml或15ml PP离心管,盖垫必须是PTFE内衬,整盒预灭菌。采购时可以要求供应商提供材质证明或耐化学性表,确认氯仿兼容。
E
Extract 2026-08-11
氯仿除脂前处理最好在通风橱内完成,避免吸入和皮肤接触。离心时如果管盖垫已经变形,说明溶剂蒸气已经渗入,不要再继续用,否则高速离心可能漏液污染离心机转子。选离心管时做个小测试:取几个不同材质盖垫,放入少量氯仿密封,室温放置24小时,观察垫片是否发粘、膨胀、漏液。这个方法比只看品牌宣传更可靠。操作上不要装太满,血清和氯仿混合后留1/3以上空间,防止涡旋时压力顶开盖子。
V
Vivi1307 2026-08-20
你这种用量不大但频次高,不建议买散装或大包装离心管,因为开盖后反复取用容易污染,盖垫也容易丢失。划算的做法是固定两个规格:1.5/2ml用于涡旋、离心小体积;15ml用于血清加氯仿除脂和转移。买整箱的预灭菌PP离心管,配PTFE复合盖垫,单价可能略高但漏液率低,实际成本更省。氯仿取液最好不用普通移液枪,用带PTFE活塞的瓶口分液器或玻璃注射器,分液器前期投入高但长期用量小,枪头损耗会明显下降。
Write an AnswerSign in to answer
Sign in required to answer (browsing is open). Sign in
✉Email 💬Message 👤Member